Next > < Prev Servicing and Repairs

Gas Bench

Once the quote is returned to LANDTEC, the instrument is serviced. Below are some typical services:

Internal filters changed
All internal filters are replaced.
Oxygen (O2) cell re-calibrated
The oxygen sensor will be re-calibrated, as the sensor output changes during its life-span this is usually carried out on all analyzers. This includes checking the 'RAW' value output of the sensor at various oxygen concentrations and with the use of proprietary software the instrument is programmed with the oxygen sensor characteristics.
Barometric sensor re-calibrated
The barometric sensor will be re-calibrated, as the sensor output may change during its life-span. This includes checking the 'RAW' value output of the sensor at known pressure levels and with the use of proprietary software the instrument is programmed with the sensor characteristics.
Methane (CH4) and Carbon Dioxide (CO2) sensors re-calibrated
The CH4 and CO2 sensors will be re-calibrated with the use of a temperature controlled environment and sophisticated computer analysis techniques. Click here to for more details on calibration.
All necessary modifications performed
The gas analyzer will automatically be upgraded with any necessary developments/modifications introduced in the period from the last service return.
Approved repairs
Any repairs quoted and accepted by the customer will be carried out.

Once servicing, repairs and calibrations are completed, the instrument goes through a Quality Control/Quality Assurance check.