Test Board

After servicing, repair and calibration, the unit receives a final Quality Control/Quality Assurance check:

General operations
The main functions of the gas analyzer's operation are checked to ensure that they are within specification.
Reading storage/download
The instrument data storage operation is fully checked, including ID entry and data download/upload.
Barometric pressure reading
The barometric pressure reading is checked to ensure it is within specification. This is carried out by comparing the atmospheric reading against a known standard.
Static and differential pressure readings
The static and differential pressure transducers are checked by applying a known pressure and noting both readings.
Flow fail setting
The flow fail function is checked to ensure proper operation with the specified limits.
Gas pod and Temperature probe connectivity reading
The connectivity of the gas analyzer is checked with all necessary accessories to ensure correct operation and reading performance.
Computer controlled gas check
Outward gas checks are carried out by way of connecting the gas analyzer to a custom built computer controlled 'gas checking rig' and proprietary software. During this process a range of gases are used that span the reading range of the gas analyzer.
Battery capacity graph display
The battery capacity graph setting is checked by way of monitoring the battery voltage.
Pump Performance
The performance of the gas analyzer's internal sample pump is checked to ensure flow rate and vacuum is within specification.

The unit is not returned to the customer until it passes final inspection.