GEM-2000 Easy Steps - Field Calibration
Printable Version
Field Calibrating the GEM-2000
Note: Be sure to always have the GEM-2000 instrument turned on prior to connecting the calibration gas bottles.
Turn Instrument On (By pressing on the Red Button).
After the instrument has finished it's initial process, it will enter the Main Screen.
Press 1 Menu
Use your Up(2) or Down(8) keys to scroll the "Field Calibration" then press Enter in your instrument
Press 3 Edit target Concentrations
Enter manually the target concentrations for CH
4
Press Enter to move to CO
2
, Enter CO
2
Target Concentration
Press Enter to move to O
2
, Enter O
2
Target Concentration
Press Enter to complete.
Connect Calibration (CAL) Gas Bottle with 0% CH
4
(example 4% O
2
, Bal N
2
)
Allow gas to flow for approximately 30 seconds to ensure a compete purge of any gas in instrument
Press Enter to bring up Calibration Menu
Choose Zero Channels, Press Enter
Choose Zero CH
4
Press Enter when done.
If Calibration was successful you should see the following message “User Zero Complete”
Connect CAL Gas Bottle With 0% O
2
(example 50% CH
4
35%CO
2
Bal N
2
)
Allow gas to flow for approximately 30 seconds to ensure a compete purge of any gas in instrument
Press Enter to bring up Calibration Menu
Choose Zero Channels press Enter
Choose Zero O
2
press Enter when done
If Calibration was successful you should see the following message “User Zero Complete”
Press Enter to go to Calibration Menu
Choose Span Channels, Press Enter
Choose Span CH
4
Verify that CH
4
Calibration (CAL) Gas is still connected to instrument (Wait 30 seconds for gas to flow thru instrument)
Press Enter to Set CH
4
Span
The following message should appear “Calibration Complete”
Press Enter to go to Calibration Menu
Choose Span Channels, Press Enter
Choose Span CO
2
Verify that CH
4
CAL Gas is still connected to instrument (Wait 30 seconds for gas to flow thru instrument)
Press Enter to Set CO
2
Span
The following message should appear “Calibration Complete”
Connect Gas Bottle With 4% O
2
(example. 0% CH
4
0% CO
2
4%O
2
)
Allow gas to flow for approximately 30 seconds to ensure a compete purge of any gas in instrument
Press Enter to go to Calibration Menu
Choose Span Channels, Press Enter
Choose Span O
2
Press Enter to Set O
2
Span
“Field Calibration is Complete”
We suggest taking a reading from CAL GAS Bottles to verify that calibration has been done correctly.
Support
MSDS Sheets
Manuals
TechTips
FAQ
Troubleshooting
Check RA Status
Easy Steps
GEM™-2000
GA-Mode
GEM™-500
Getting Online
Field Calibration
Servicing
Return Authorization
Shipping
Incoming Assessment
Quotes
Service/Repair
Outgoing QC
Return Shipping
Calibration Details
Calibration Certificate
Home
|
AEMS
|
Products
|
Support
|
News
|
About
|
Contact Us
Copyright © 2007 LANDTEC North America